The meaning of free space (dead volume) is that it is not the geometrical volume of the system but it is the volume of convenience used for the calculation of adsorption amount. In case of nitrogen adsorption measurement, the sample is cooled down by liquid nitrogen. After cooling, there will be two different temperature zones in the sample cell. The gas density of the cooled zone is higher and it is influenced by the coolant temperature and by the change in the coolant level. The volumetric method counts the number of un-adsorbed gas molecules (density), and in the process the dead volume plays an important role. The conventional volumetric measurement procedure is to cool down the sample by liquid nitrogen, measure the dead volume by non-adsorbing He gas at the liquid nitrogen temperature, and then measure the nitrogen or other adsorptive gas isotherm(*). During the adsorption measurement the coolant level needs to be maintained.
What's the dead volume?
The dead volume(*) is expressed by the following equation. It is the apparent volume of cooled space that is converted from the volume at the coolant temperature to the volume at the apparatus temperature.
On the other hand, our new method, AFSMTM, can continuously measure the change of dead volume at each data point without maintaining the coolant level. This new volume calibration method (AFSMTM) has introduced us more accurate adsorption measurement.
Contactez nous pour un devis gratuit
En fin de compte, le choix d'utiliser une solution de tamisage simple ou d'investir dans la diffraction laser ou l'analyse dynamique d'images dépendra du volume d'essais, du budget et du personnel disponibles, ainsi que des normes internationales spécifiques ou des exigences des clients auxquelles vous devez faire face.
Pourquoi ne pas contacter Microtrac pour une consultation gratuite afin de déterminer quelle solution vous apportera les résultats et le retour sur investissement dont vous avez besoin ?